Masimo® Plethysmograph Variability Index as a Tool for Assessment of Fluid Responsiveness in Elective Major Abdominal Surgeries
Ahmed M. Essam, Mohamed Z. Ali, Mohamed A. Maher, Ali M. Mokhtar, Sohila H. Omar, Hossam H. El-Sabae, Mohamed H. Hafez
Journal of Advances in Medicine and Medical Research · pp. 1–11 · Published 31 May 2016
10.9734/BJMMR/2016/26030Abstract
Background: Maximizing the stroke volume (SV) as measured by Trans-oesophageal Doppler (TED) optimizes preload, & is a goal-directed fluid therapy technique that has been used in a variety of clinical settings. Masimo® Plethysmograph variability Index (PVI) is a reliable, safe & noninvasive tool to guide fluid management. PVI is an automated measure of the dynamic change in the perfusion index (PI) that occurs during a respiratory cycle. This study was designed to determine whether PVI, measured using finger co-oximetry is an efficient predictor of fluid responsiveness in low-risk patients undergoing elective major abdominal surgery. Subjects and Methods: 60 ASA I-II patients of either sex, 25-60 years old, undergoing major abdominal surgery were enrolled in this study. A Masimo® Radical-7 Pulse Co-Oximeter probe & a Cardio Q TED probe were applied to each patient. In all patients, a fluid bolus of 500 ml of 130/0.4 tetrastarch colloid solution was administered rapidly via pressurized IV infusion. Maintenance & deficits were calculated routinely. If the SV decreased by 10%, a 250-mL bolus of colloid was given via fast infusion. Patients’ demography, TED-derived measurements: (SV & Flow Time corrected (FTc)), Masimo®-derived measurements: (PVI & PI), HR and MAP were all collected and statistically analyzed. Measurements were done at five minutes post-induction T1, Ten minutes after volume expansion (500 ml colloid) T2, If the SV decreased by 10%, (guided by TED) T3, Then 250 ml colloid is given. Ten minutes after a 250-ml colloid bolus T4. Results: A significant difference was found in FTc, SV, PI & PVI in T1 vs. T2 & T3 vs. T4 (P=0.001). There was a significant difference in PI & PVI between responders & non-responders for the 1st bolus (P<0.05) and in SV & PVI in subsequent boluses (P<0.01). There was no significant difference between percent changes of SV and PVI at T3 & T4. Conclusions: Plethysmograph Variability Index (PVI) measured by Masimo® Co-Oximeter is an efficient predictor of fluid responsiveness as SV measured by TED in low risk patients undergoing elective major surgery.
References (15)
- 1 A method of comparing the areas under receiver operating characteristic curves derived from the same cases. [DOI]
- 2 Relation between Respiratory Changes in Arterial Pulse Pressure and Fluid Responsiveness in Septic Patients with Acute Circulatory Failure [DOI]
- 3 Intraoperative fluid optimization using stroke volume variation in high risk surgical patients: results of prospective randomized study [DOI]
- 4 Goal-directed fluid management based on pulse pressure variation monitoring during high-risk surgery: a pilot randomized controlled trial [DOI]
- 5 Pleth variability index to monitor the respiratory variations in the pulse oximeter plethysmographic waveform amplitude and predict fluid responsiveness in the operating theatre [DOI]
- 6 Randomised controlled trial investigating the influence of intravenous fluid titration using oesophageal Doppler monitoring during bowel surgery* [DOI]
- 7 Accuracy of stroke volume variation compared with pleth variability index to predict fluid responsiveness in mechanically ventilated patients undergoing major surgery [DOI]
- 8 Meta-analysis of hemodynamic optimization: relationship to methodological quality [DOI]
- 9 The Ability of a Novel Algorithm for Automatic Estimation of the Respiratory Variations in Arterial Pulse Pressure to Monitor Fluid Responsiveness in the Operating Room [DOI]
- 10 Physicians’ Estimates of Cardiac Index and Intravascular Volume Based on Clinical Assessment Versus Transesophageal Doppler Measurements Obtained by Critical Care Nurses [DOI]
- 11 A comparison of hemodynamic data derived by pulmonary artery flotation catheter and the esophageal Doppler monitor in preeclampsia [DOI]
- 12 Dynamic variables and fluid responsiveness in patients for aortic stenosis surgery [DOI]
- 13 Pleth Variability Index and Fluid Responsiveness of Hemodynamically Stable Patients After Cardiothoracic Surgery [DOI]
- 14 Clinical Evaluation of the HemoSonic Monitor in Cardiac Surgical Patients in the ICU [DOI]
- 15 Incorrectly placed oesophageal Doppler probe [DOI]
Cited by 1
Article metrics
Real usage data collected on this platform.
7
Page views
0
PDF downloads
8
Outbound clicks
1
Citations
Views over time
Views by country
Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".
Traffic sources
Referring site, by host.
Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.