Qualitative Assessment of X-Ray Machines’ Mechanical Parameters at Some Selected Diagnostic Centres in Nigeria
Owoade Latifat R., Akinlade Bidemi I., Farai Idowu P.
Journal of Advances in Medicine and Medical Research · pp. 436–443 · Published 30 Aug 2022
10.9734/jammr/2022/v34i2131562Abstract
Introduction: This study assessed the present status of conventional X-Ray machines used for common diagnostic X-Ray examinations in Nigeria. Materials and Methods: Total of 112 X-Ray machines from 106 diagnostic centres from six geopolitical zones in Nigeria were considered. Structured questionnaires and Diavolt Dosimeter were respectively used for data collection and measurements of parameters: Half-Value Layer (HVL), tube voltage (kVp) reproducibility, kVp accuracy and beam output linearity/stability. Results: By zonal distribution of X-Ray units, the highest number (36%) and the least (7%) was from the south-west and south-south respectively. In all, 71% were owned by private institutions, 71% were non-mobile, 64% were operated on 3-phase generator and 35% were within the age of ten years. The HVLs of 97% of the X-Ray units were within the standard thickness of 2.3 mmAl (at 80 kVp, 100 cm FFD), 88% have acceptable kVp reproducibility and 51% have kVp accuracy within ±5%. Only X-Ray units (64%) operated on 3-phase generator produced beam output that are within acceptable stability. Conclusion: This study showed that some (65%) of the X-Ray units considered are aged (>10 years) and about 49% of the units failed kVp accuracy test. This may be due to inconsistency in the quality assurance checks of X-Ray units by some diagnostic centres, especially those in the south-south region. The authors therefore recommended strict measures, to be put in place by the regulatory authority, that will make all diagnostic centres in Nigeria to comply with the requirements for peaceful radiation practices.
Cited by 0
No indexed citations yet.
Related research
- Effect of Slice Thickness Variation on Spatial Resolution of CT-Scan Device Images at the Bali Mandara Hospital Radiology Installation — shares topic coverage
- Impact of Body Mass Index and Sex on Image Quality and Radiation Dose of Routine CT Scan — shares topic coverage
- Optimizing the Dose-Quality Paradigm in Low-Dose CT Lung Cancer Screening: A Narrative Review — shares topic coverage
- Effect of Contrast and Brightness on Digital Images under Outdoor Conditions — shares topic coverage
- Image Content Enhancement of Natural Images Using Genetic Algorithm — shares topic coverage
Article metrics
Real usage data collected on this platform.
32
Page views
0
PDF downloads
3
Outbound clicks
0
Citations
Views over time
Views by country
Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".
Traffic sources
Referring site, by host.
Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.