Skip to content
Research Article Open access CC BY 4.0

Interference by Human Anti-mouse Antibodies in Immunoassays: Falsely Elevated Cardiac Troponins Leading to Negative Coronary Angiograms

Mian Bilal Alam, Surachit Kumar, Ibrahim Sbietan, K. M. Anwar Hussain, Mehwish Asad, Asif N. Khan

Annual Research & Review in Biology · pp. 323–327 · Published 19 Feb 2015

10.9734/ARRB/2015/11941

Abstract

Cardiac Troponins are an extremely important indicator of myocardial injury and American College of Cardiology (ACC) and American Heart Association (AHA) guidelines include Cardiac Troponins as one of the major diagnostic criteria for myocardial damage. However in some instances the presence of heterophilic antibodies like Human Anti-Mouse Antibodies (HAMA) may interfere with the sandwich assays used for the detection and quantification of cardiac troponins giving false results which can have a major impact on the management of an individual suspected to have myocardial injury. We present a case of a 56 year old Caucasian male who had negative coronary angiogram 2 years ago presenting with acute chest pain and significantly elevated levels of Cardiac Troponin I (cTnI). He underwent cardiac catheterization which again turned out to be completely normal. His cTnI levels continued to be consistently elevated for months after cardiac catheterization. The reason for elevated cTnI levels was due to high levels of HAMA antibodies that caused heterophilic antibody interference in the assay of cTnI, resulting in falsely elevated levels of cTnI.

Cardiac troponin I (cTnI) cardiac troponin T (cTnT) human anti-mouse antibodies (HAMA) heterophilic antibodies

References (22)

  1. 1 2012 ACCF/AHA Focused Update of the Guideline for the Management of Patients With Unstable Angina/Non–ST-Elevation Myocardial Infarction (Updating the 2007 Guideline and Replacing the 2011 Focused Update) [DOI]
  2. 2 Human Anti-Animal Antibody Interferences in Immunological Assays [DOI]
  3. 3 Heterophilic antibodies: a problem for all immunoassays. [DOI]
  4. 4 Incidence and specificity of interference in two-site immunoassays. [DOI]
  5. 5 False Increase of Cardiac Troponin I with Heterophilic Antibodies [DOI]
  6. 6 Human Anti-Mouse Antibodies [DOI]
  7. 7 Interference from heterophilic antibodies in troponin testing. Case report and systematic review of the literature [DOI]
  8. 8 Falsely Elevated Cardiac Troponin I Levels [DOI]
  9. 9 Radioimmunotherapy: recent results and future directions. [DOI]
  10. 10 Cardiac troponin I to diagnose myocardial injury [DOI]
  11. 11 Circulating Antibodies to Bovine Albumin in Ulcerative Colitis and Crohn's Disease [DOI]
  12. 12 Transient Neonatal Hyperthyrotropinemia: A Factitious Syndrome due to the Presence of Heterophilic Antibodies in the Plasma of Infants and Their Mothers [DOI]
  13. 13 Heterophilic Antibodies Against Rabbit Serum Causing Falsely Elevated Gonadotropin Levels [DOI]
  14. 14 Population Study of Heterophile Antibodies [DOI]
  15. 15 Unnecessary coronary angiography due to false positive troponin I results in a 51-year-old man [DOI]
  16. 16 Clinical Applications of Monoclonal Antibody Technology [DOI]
  17. 17 Interference by heterophilic antibodies in immunoassays: wrong increase of myoglobin values.
  18. 18 False-positive troponin I in a young healthy woman with chest pain.
  19. 19 Anti-Antiserum Antibody as a Cause of Double Precipitin Rings in Immunoglobulin Quantitation and Its Relation to Milk Precipitins [DOI]
  20. 20 The use of antibodies in clinical cardiology [DOI]
  21. 21 Interference by Human Antibodies with Tumor Marker Assays [DOI]
  22. 22 [Two cases of false troponin I increase in patients with heterophile antibodies].

Cited by 0

No indexed citations yet.

Article metrics

Real usage data collected on this platform.

4

Page views

0

PDF downloads

7

Outbound clicks

0

Citations

Views over time

Views by country

Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".

Traffic sources

Referring site, by host.

Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.